Background and Purpose: Brainomix 360 Stroke is an artificial intelligence-based decision support tool that aids the interpretation of CT imaging in the context of acute stroke. While Brainomix 360 Stroke has the potential to improve the speed and accuracy of diagnosis, real-world validation is essential. The aim of this study was to prospectively evaluate the performance of Brainomix 360 Stroke in an unselected cohort of patients with suspected acute ischemic stroke.

Methods: The study cohort included all patients admitted to the University College London Hospital Hyperacute Stroke Unit between October 2021 and April 2022. For e-ASPECTS and e-CTA, the ground truth was determined by a neuroradiologist with access to all clinical and imaging data. For e-CTP, the values of the core infarct and ischemic penumbra were compared with those derived from syngo.via, an alternate software used at our institution.

Results: A total of 1,163 studies were performed in 551 patients admitted during the study period. Of these, 1,130 (97.2%) were successfully processed by Brainomix 360 Stroke in an average of 4 minutes. For identifying acute middle cerebral artery territory ischemia, e-ASPECTS had an accuracy of 77.0% and was more specific (83.5%) than sensitive (58.6%). The accuracy for identifying hyperdense thrombus was lower (69.1%), mainly due to false positives (positive predictive value of 22.9%). Identification of acute hemorrhage was highly accurate (97.8%) with a sensitivity of 100% and a specificity of 97.6%; false positives were typically caused by areas of calcification. The accuracy of e-CTA for large vessel occlusions was 91.5%. The core infarct and ischemic penumbra volumes provided by e-CTP strongly correlated with those provided by syngo.via (ρ = 0.804–0.979).

Conclusion: Brainomix 360 Stroke software provides rapid and reliable analysis of CT imaging in the acute stroke setting. However, in line with the manufacturer's guidance, it should be used as an adjunct to expert interpretation rather than a standalone decision-making tool.

Our Partners

Leave Your Legacy Portuguese Logo
Stryker
OSIC Logo
Boehringer Ingelheim Logo
Nvidia Logo
NCIMI Logo
GE Healthcare
Wallaby Logo
TCPDS MEMBER BLACK
Oxlep ERDF BW
Blackford
Fora Logo
Calantic Digital Solutions Logo